Home chromatograph Gas Chromatograph Portable Photo-Ionization Gas Chromatograph GC-4400

Portable Photo-Ionization Gas Chromatograph GC-4400



Portable Photo-Ionization Gas Chromatograph GC-4400 uses long life-span vacuum UV lamp and has high sensitivity, low limit of detection with broad linear range. It does not need H2, nor the combustion-assistant gases. It can perform trace analysis not only for organic compounds but also for some inorganic compounds. Due to the high sensitivity, when analyzing pollutant from indoor or outdoor environ-ment, there is no need of thermo desorption devices for sample enrichment. Portable Photo-Ionization Gas Chromatograph has wide applications in fields of environment protection, indoor air quality control, food safety, national defense, anti-chemical war, anti-terrorism war, aviation, aerospace, scientific research, and others requiring trace gas (vapor) analysis.

Portable Photo-Ionization Gas Chromatograph GC-4400


Technical Specifications:

Utilize photon to realize ionization of samples. No radiation source.
High sensitivity. Low limit of detection. Broad linear range.
Light weight, manoeuvrable and flexible, especially suitable for use in analysis at scene and emergency for environment pollution accidents.
No need of H2, combustion-assistant gases and high purity gas. Only mixed air gas is needed.
No destrution of samples. The results are safe and reliable.
No need of chemical pre-processing, concentration, enrichment, thermo desorption. The user can directly inject samples to analyze.
Chromatograph workstation, composed of data acquisition card, a notebook computer and software. It carries out of data sampling, analysis, calculation, display and storing.

Principle of Vacuum UV Light Source and Photo-Ionization:



Limit of detection: 0.3ppb (Benzene)
Lenear scope: 105
Photon energy: 10.6 eV
Life of light source: ≥3000 hours
Weight: 14 kg

Compounds for Analysis:

1ppb: Epoxy-ethane, ethene-chloride, ethane-chloride, benzene, toluene, xylene, ethane, paraffin hydrocarbon (up to octyl-alkane), isoamyl diene, ethane, propylene, butylenes, category of light weight chlorobenzene, H2S, category of light weight mercaptan, organic sulfur (up to dimenthyl disulfide), acetone, methyl-ethyl-ketone, arsine, hydrogen phosphide, acetaldehyde, category of aldehyde (up to hexyl-aldehyde) etc.
50ppb: Category of ethylene lipid, category of fluoro-chloro-methane (freon), isocyanic-acid-methyl-lipid, category of ethyl-chloride, cyclo-hexyl-ketone, acryclic-acid-ethyl-lipid, category of light weight alcohol and etc.
Repeated Measurement of Benzene: Quantitative Method: Standard Curve with Added Standard (Peak Height) Conditions: Temperature 31℃, Carrier Gas N2 15mL/min Sample: 0.1mL 100ppb Benzene Column for Benzene: 10% SE-30, 1.2m×φ3mm PTFE Packed Column